"Application of the HeartLander crawling robot for injection of a thermally sensitive anti-remodeling agent for myocardial infarction therapy"

Publication information:

Chapman MP ", Gonzalez JLL, Goyette B.
"Application of the HeartLander crawling robot for injection of a thermally sensitive anti-remodeling agent for myocardial infarction therapy". "Annu. Int. Conf. IEEE Eng. Med. Biol. Soc.". 2010;2010:"5428–5431".

Abstract

"The injection of a mechanical bulking agent into the left ventricular (LV) wall of the heart has shown promise as a therapy for maladaptive remodeling of the myocardium after myocardial infarct (MI). The HeartLander robotic crawler presented itself as an ideal vehicle for minimally-invasive, highly accurate epicardial injection of such an agent. Use of the optimal bulking agent, a thermosetting hydrogel developed by our group, presents a number of engineering obstacles, including cooling of the miniaturized injection system while the robot is navigating in the warm environment of a living patient. We present herein a demonstration of an integrated miniature cooling and injection system in the HeartLander crawling robot, that is fully biocompatible and capable of multiple injections of a thermosetting hydrogel into dense animal tissue while the entire system is immersed in a 37°C water bath."